ProLab 4000
pH
8
3
ba75646e03
12/2012
7.3.2
Carrying out a calibration procedure (example:
VariCal
)
Make sure the
Calibration
/
pH calibration type
/
VariCal
calibration type
is marked (see page 73).
Single-point calibration
Use any buffer solution for this rapid method.
The calibration will be the more exact the nearer the pH value of the
buffer solution is to that of the test sample.
Two-point calibration
Use two buffer solutions for this procedure:
first buffer solution: pH 7.000 ± 0.500
any other buffer solution
To carry out the calibration procedure, the
pH calibration type
VariCal
must be set in the sensor menu (see page 87).
1
Perform the preparatory activities (see page 78).
2
Start the calibration with
<CAL>
.
The
pHX calibration
window opens.
3
When measuring without temperature sensor:
Measure the temperature of the standard solution using a
thermometer.
Open the setting window for the temperature value with
[Man. temp.]
.
Enter the temperature and confirm with
[OK]
.
4
Thoroughly rinse the electrode with distilled water.
5
Immerse the electrode in buffer solution 1 (pH 7.000 ± 0.500).
6
Start the measurement with
[Continue]
.
The measured value is checked for stability (Stability control).
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......