ProLab 4000
Ion concentration
115
ba75646e03
12/2012
6
In the
Sample
register, enter the values for the sample volume
(
Sample volume
) and the volume of the ISA/TISAB solution
(
ISA/TISAB volume
) as necessary.
7
Use
[Dosing]
to automatically add the adjusted volume of the
ISA/TISAB solution (
ISA/TISAB volume
).
The
[Continue]
button appears after the dosing.
or
If the burette control is inactive:
Add the adjusted volume of the ISA/TISAB solution (
ISA/TISAB
volume
) manually.
8
Thoroughly rinse the electrode with deionized water.
9
Immerse the electrode in the sample.
10
Start the measurement with
[Continue]
.
The measured value is checked for stability (Stability control).
11
Wait for the measurement with stability control to end.
or
Take over the calibration value with
[OK]
.
The
Standard
register with the calculated measured value
pops up.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......