11.1.1 Creating and selecting a file for manually
stored measurement datasets
Creating a file for
manually stored
measurement datasets
You can create the file in which to store your manually stored measure-
ment datasets.
Files for manually stored measurement datasets have the file exten-
sion, "*.mst".
Selecting the file for
measurement dataset to
be stored manually
You can open an existing file for stored measurement datasets in order
to append new measurement datasets.
Files for manually stored measurement datasets have the file exten-
sion, "*.mst".
1
Open the menu,
Memory
/
Manual memory
/
New...
.
The
Open file
dialog box pops up.
2
Enter a new file name and confirm with
[OK]
.
The file is created. Data from manual storing will be stored in it.
Note
If a user has not yet created a file, the data is stored in the file, "Man-
Sto.mst".
1
Open the menu,
Memory
/
Manual memory
/
Select...
.
The
Open file
dialog box pops up.
2
Select a file and confirm with
[OK]
.
Manually stored data is written in the file.
New datasets are appended to the existing dataset.
Содержание ProLab 4000
Страница 4: ...ProLab 4000 4 ba75646e03 12 2012...
Страница 10: ...Contents ProLab 4000 10 ba75646e03 12 2012...
Страница 20: ...Overview ProLab 4000 20 ba75646e03 12 2012...
Страница 30: ...Commissioning ProLab 4000 30 ba75646e03 12 2012...
Страница 68: ...System functions ProLab 4000 68 ba75646e03 12 2012...
Страница 86: ...pH ProLab 4000 86 ba75646e03 12 2012 Note A calibration line is determined for the two point calibration...
Страница 138: ...Ion concentration ProLab 4000 138 ba75646e03 12 2012...
Страница 184: ...Recorder ProLab 4000 184 ba75646e03 12 2012...
Страница 202: ...What to do if ProLab 4000 202 ba75646e03 12 2012...
Страница 216: ...Lists ProLab 4000 216 ba75646e03 12 2012...
Страница 227: ......