Remote Commands
R&S
®
Spectrum Rider FPH
314
User Manual 1321.1011.02 ─ 06.00
CALCulate<1...2>:MARKer<1...6>:MAXimum[:PEAK]
This command positions a marker on the current trace maximum.
If necessary, the corresponding marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:MAX
Moves marker 2 to the maximum peak.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:MAXimum:NEXT
This command positions a marker on the next smaller trace maximum.
If necessary, the corresponding marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:MAX:NEXT
Moves marker 2 to the next smaller maximum peak.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:MINimum[:PEAK]
This command positions a marker on the current trace minimum.
If necessary, the corresponding marker is activated first.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Example:
CALC:MARK2:MIN
Moves marker 2 to the trace minimum.
Usage:
Event
CALCulate<1...2>:MARKer<1...6>:X
<FrequencyOrTime>
This command positions a marker on a particular coordinate on the horizontal axis.
If one or more delta markers (2 to 6) are already active, the command turns these delta
markers into normal markers.
Note that it is possible to place the marker outside the visible trace. In that case, this
value is invalid.
Remote Control - Commands