The configuration for a standard is in line with the specifications and includes parame-
ters like weighting filter, channel bandwidth and spacing, resolution and video band-
width, as well as detector and sweep time.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<Standard>
String containing the filename of the standard.
Example:
CALC:MARK:FUNC:POW:PRES '3GPP WCDMA.chpstd'
Selects the 3GPP WCDMA standard for channel power mea-
surements.
Usage:
Setting only
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:PRESet:CHECk?
This command checks if the current settings are in line with the settings defined by the
standard selected with
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Note that the command only checks those parameters that are defined by the stan-
dard.
This command is a query and therefore has no *RST value.
Return Value
0
The R&S Spectrum Rider settings violate a standard
configuration
1
The R&S Spectrum Rider settings comply with the
standard.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Example:
CALC:MARK:FUNC:POW:PRES:CHEC?
Queries compliance to the standard currently in use.
Usage:
Query only
CALCulate<1...2>:MARKer<1...6>:FUNCtion:LEVel:ONCE
This command adjusts the reference level to the measured signal power.
This automatic routine makes sure that the signal power level does not overload the
R&S Spectrum Rider or limits the dynamic range by too small an S/N ratio.
To determine the best reference level, the R&S Spectrum Rider aborts current mea-
surements and performs a series of test sweeps. After it has finished the test, it contin-
ues with the actual measurement.
This command is an event and therefore has no *RST value and no query.