Remote Commands
R&S
®
Spectrum Rider FPH
295
User Manual 1321.1011.02 ─ 06.00
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLvlOffset>
Sets the reference level offset.
Range:
-100 dB to 100 dB
*RST:
0 dB
Default unit: dB
Example:
DISP:TRAC:Y:RLEV:OFFS –10dB
DISPlay<1...2>[:WINDow]:TRACe<1...2>:Y[:SCALe]:RPOSition
<RefLvlPosition>
This command defines the position of the reference level on the display grid.
First, you have to set a logarithmic scale for the vertical axis with
WINDow]:TRACe<1...2>:Y:SPACing
on page 293. For a linear scale, you cannot
modify the reference position.
The numeric suffix at DISPlay and TRACe are irrelevant for this command.
Parameters:
<RefLvlPosition>
Defines the reference position in %.
Range:
0 to 10
*RST:
10
Example:
DISP:TRAC:Y:RPOS 5
Sets the reference position to the 5th grid line.
INPut:ATTenuation
<Attenuation>
This command defines the input attenuation.
The attenuation is coupled to the reference level. If you set the attenuation independ-
ently, the R&S Spectrum Rider turns off this coupling.
The R&S Spectrum Rider adjusts the reference level if it cannot be set for the current
RF attenuation.
Parameters:
<Attenuation>
Range:
0 dB to 40 dB
*RST:
0 dB (AUTO is ON)
Default unit: dB
Example:
INP:ATT 30dB
Defines an attenuation of 30 dB and deactivates coupling to the
reference level.
INPut:ATTenuation:MODE
<AttenMode>
This command selects the attenuation mode.
Remote Control - Commands