Remote Commands
R&S
®
Spectrum Rider FPH
315
User Manual 1321.1011.02 ─ 06.00
If necessary, the corresponding delta marker is activated first.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<FrequencyOrTime> Indicates the coordinate on the horizontal axis.
The unit in spectrum analyzer mode depends on the measure-
ment, e.g. Hz for measurements in the frequency domain and
seconds for measurements in the time domain.
Range:
Maximum span.
Default unit: GHz
Example:
CALC:MARK2:X 10.7MHz
Positions marker 2 to frequency 10.7 MHz.
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>[:STATe]
<State>
This command turns marker search limits on and off.
The search limit restricts the evaluation range of the trace when "set marker" functions
are performed. For example, marker set to peak with command
on page 314 sets the marker only at the peak of
the trace within the search limit.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
See
CALCulate<1...2>:MARKer<1...6>:X:
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>:LEFT
<SearchLimit>
This command defines the left limit of the marker search range.
To use the command, you first have to turn on search limits with
CALCulate<1...2>:MARKer<1...6>:X:SLIMits<1...2>[:STATe]
on page 315.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MARKer selects the marker.
Remote Control - Commands