Remote Commands
R&S
®
Spectrum Rider FPH
353
User Manual 1321.1011.02 ─ 06.00
Usage:
Event
Measuring the AM Modulation Depth
The following commands configure AM Modulation Depth measurements. To perform
the actual measurement, use the commands described in section
"Performing and Triggering Measurements"
List of commands
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth[:STATe]
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth:RESult?
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth[:STATe]
<State>
This command turns the AM Modulation Depth measurement on and off.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MARK:FUNC:MDEP ON
Activates the AM modulation depth measurement.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:MDEPth:RESult?
This command queries the results of the AM modulation depth measurement.
The suffix at CALCulate and MARKer are irrelevant for this command.
Parameters:
<Return values>
<ModulationDepth>
Modulation depth in %.
Example:
CALC:MARK:FUNC:MDEP:RES?
Queries the AM modulation depth.
Usage:
Query only
Measuring the Spectrum Emission Mask
The following commands configure Spectrum Emission Mask (SEM) measurements..
To perform the actual measurement, use the commands described in section
ter 13.6.2.4, "Performing and Triggering Measurements"
List of commands
●
●
[SENSe:]ESPectrum:PRESet[:STANdard]
Remote Control - Commands