Remote Commands
R&S
®
Spectrum Rider FPH
336
User Manual 1321.1011.02 ─ 06.00
Measuring the Channel Power
The following commands configure channel power measurements. To perform the
actual measurement, use the commands described in section
forming and Triggering Measurements"
List of commands
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CPOWer:BANDwidth
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CPOWer:MODE
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CPOWer:UNIT
●
CALCulate<1...2>:MARKer<1...6>:FUNCtion:POWer:RESult:PHZ
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CPOWer:BANDwidth
<ChannelBW>
This command defines the channel bandwidth for channel power measurements.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<ChannelBW>
*RST:
3.84 MHz
Default unit: GHz
Example:
CALC:MARK:FUNC:CPOW:BAND 4 MHZ
Sets the channel bandwidth to 4 MHz.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CPOWer:MODE
<DisplayMode>
This command selects the display mode for channel power measurements.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Parameters:
<DisplayMode>
CLR | MAX
CLR
Displays the currently measured value (Clear / Write).
MAX
Displays the highest measured value (Max Hold).
*RST:
CLR
Example:
CALC:MARK:FUNC:CPOW:MODE CLR
Selects clear/write power display mode.
CALCulate<1...2>:MARKer<1...6>:FUNCtion:CPOWer:UNIT
<Unit>
This command selects the unit of the vertical axis for channel power measurements.
The numeric suffix at CALculate and MARKer are irrelevant for this command.
Remote Control - Commands