Remote Commands
R&S
®
Spectrum Rider FPH
360
User Manual 1321.1011.02 ─ 06.00
CALCulate<1...2>:ADEModulation:LIMit:SELect
<Filename>
This command loads a limit line for the selected measurement mode.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<Filename>
String containing the filename of the limit line.
Example:
CALC:ADEM:LIM:SEL 'GSM.aamlim'
Selects the filename for limit line.
CALCulate<1...2>:ADEModulation:LIMit:DELete
This command clears the selected limit.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Example:
CALC:ADEM:LIM:DEL
Clears off the selected limit line.
Usage:
Event
CALCulate<1...2>:ADEModulation:LIMit:FAIL?
This command queries the result of a limit check.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end before reading out the result. This is only possible in single sweep mode.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Example:
CALC:ADEM:LIM:FAIL?
Queries the result of the check for limit line
Usage:
Query only
13.6.3.7
Setting the Measurement Mode
The following chapter describes command necessary to switch to the correct measure-
ment mode.
List of commands
●
●
Remote Control - Commands