Remote Commands
R&S
®
Spectrum Rider FPH
367
User Manual 1321.1011.02 ─ 06.00
13.6.4.3
Defining the Measurement Time
The following chapter describes commands to define the measurement time of the
power sensor.
List of commands
●
[SENSe:]PMETer:MTIMe
<MeasTime>
This command sets the duration of measurements.
Available for measurements with a power sensor.
Parameters:
<MeasTime>
SHORt | NORMal | LONG
Example:
PMET:MTIMe SHOR
Sets a short measurement time for power measurements.
13.6.4.4
Zeroing of the Power Sensor
List of commands
●
CALibration<1...2>:PMETer:ZERO:AUTO
CALibration<1...2>:PMETer:ZERO:AUTO
<ONCE>
This command starts to zero the power sensor.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALibration is irrelevant for this command.
Parameters:
<ONCE>
ONCE
Example:
CAL:PMET:ZERO:AUTO ONCE
Starts to zero the power meter.
13.6.4.5
Forward Power Display
Note that the forward power is only available in conjunction with the R&S FSH-Z14 or
R&S FSH-Z44.
List of commands
●
[SENSe:]PMETer:DETector[:FUNCtion]
[SENSe:]PMETer:DETector[:FUNCtion]
<Detector>
This command selects the forward power display of the power sensor.