STATus:PRESet
This command resets the edge detectors and ENABle parts of all registers to a defined
value. All PTRansition parts are set to FFFFh, i.e. all transitions from 0 to 1 are detec-
ted. All NTRansition parts are set to 0, i.e. a transition from 1 to 0 in a CONDition bit is
not detected. The ENABle part of the STATus:OPERation and STATus:QUEStionable
registers are set to 0, i.e. all events in these registers are not passed on.
Example:
STAT:PRES
Usage:
Event
STATus:QUEue[:NEXT]?
This command returns the earliest entry to the error queue and deletes it.
Positive error numbers indicate device-specific errors, negative error numbers are error
messages defined by SCPI. If the error queue is empty, the error number 0, "no error",
is returned. This command is identical with the command SYSTem:ERRor.
Example:
STAT:QUE?
Usage:
Query only
STATus:OPERation[:EVENt]?
This command reads out the EVENt section of the OPERation register.
The command at the same time deletes the contents of the EVENt section.
Usage:
Query only
STATus:OPERation:CONDition?
This command reads out the CONDition section of the OPERation register.
The command does not delete the contents of the EVENt section.
Usage:
Query only
STATus:OPERation:ENABle
<DecimalValue>
This command controls the ENABle part of the OPERation register.
The ENABle part allows true conditions in the EVENt part of the status register to be
reported in the summary bit. If a bit is 1 in the enable register and its associated event
bit transitions to true, a positive transition occurs in the summary bit reported to the
next higher level.
Parameters:
<DecimalValue>
Range:
0 to 65535
Remote Control - Commands