Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
956
User Manual 1178.8520.02 ─ 08
[SENSe:]ESPectrum<sb>:MSR:LTE:CPResent
<State>
This command defines whether an LTE FDD 1.4
MHz or 3
MHz carrier is located at the
edge of the specified RF bandwidth. In this case, the specification demands specific
limits for the SEM ranges.
This command is only available for band category 2 (see
Suffix:
<sb>
.
1 to 8
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
//Select the band category 2
ESP2:MSR:BCAT BC2
//Set the base station class to medium range
ESP2:MSR:CLAS MED
//Set the maximum output power to 10 dBm.
ESP2:MSR:MPOW 10
//Set the base station RF bandwidth to 20 MHz
ESP2:MSR:RFBW 20MHZ
//LTE present
ESP2:MSR:LTE:CPR ON
//Calculate limits for MSR SEM
ESP2:MSR:APPL
Manual operation:
See
"Carrier Adjacent to RF Bandwidth Edge"
[SENSe:]ESPectrum<sb>:MSR:MPOWer
<Power>
Defines the maximum output power of the base station.
This setting is only available for base stations with a medium range base station class
(see
[SENSe:]ESPectrum<sb>:MSR:CLASs
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
Parameters:
<Power>
Range:
0 dBm to 100 dBm
Increment: 1 dB
Default unit: dBm
Configuring and performing measurements
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...