Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1131
User Manual 1178.8520.02 ─ 08
[SENSe:]SWEep:EGATe
<State>
This command turns gated measurements on and off.
The measurement ends when a particular number of measurement points has been
recorded.
(See
[SENSe:]SWEep[:WINDow<n>]:POINts
Performing gated measurements turns the squelch off.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
SWE:EGAT ON
Switches on the gate mode.
SWE:EGAT:TYPE EDGE
Switches on the edge-triggered mode.
SWE:EGAT:HOLD 100US
Sets the gate delay to 100 µs.
SWE:EGAT:LEN 500US
Sets the gate opening time to 500 µs.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation:
See
See
[SENSe:]SWEep:EGATe:CONTinuous:PCOunt
<Amount>
Defines the number of gate periods to be measured after a single trigger event.
Parameters:
<Amount>
integer
Range:
1 to 1023
Increment: 1
*RST:
100
Example:
SWE:EGAT:CONT:PCO 50
Manual operation:
See
[SENSe:]SWEep:EGATe:CONTinuous:PLENgth
<Time>
Defines the length in seconds of a single gate period in continuous gating. The length
is determined from the beginning of one gate measurement to the beginning of the
next one.
Setting basic measurement parameters
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...