
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
950
User Manual 1178.8520.02 ─ 08
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:LIMit[:STATe]
<State>
This command selects the limit check mode for each power class.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 8
<pc>
1..n
power class
Parameters:
<State>
ABSolute | RELative | AND | OR
ABSolute
Evaluates only limit lines with absolute power values
RELative
Evaluates only limit lines with relative power values
AND
Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
OR
Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.
*RST:
REL
Example:
CALC:LIM:ESP:PCL:LIM ABS
Manual operation:
See
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum
<Level>
This command defines the upper limit of a particular power class.
Note:
●
The last power class always has an upper limit of 200
dBm.
●
The upper limit of a power class must always be the same as the lower limit of the
subsequent power class.
●
The power class must already exist (see
ESPectrum<sb>:PCLass<pc>:COUNt
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<sb>
1 to 8
<pc>
1..n
power class
Configuring and performing measurements
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...