Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1276
User Manual 1178.8520.02 ─ 08
Remote commands exclusive to storing results:
............................................................................................1276
...........................................................................................1276
.........................................................................................1277
......................................................................................1277
............................................................................. 1277
.................................................................................. 1278
FORMat:DEXPort:HEADer
<State>
If enabled, additional instrument and measurement settings are included in the header
of the export file for result data. If disabled, only the pure result data from the selected
traces and tables is exported.
See
Chapter 7.6.6, "Reference: ASCII file export format"
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Manual operation:
See
"Include Instrument & Measurement Settings"
MMEMory:STORe<n>:LIST
<FileName>
This command exports the SEM and spurious emission list evaluation to a file.
The file format is *.dat.
Secure User Mode
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 3.2.15, "Protecting data using the secure user mode"
Suffix:
<n>
.
Parameters:
<FileName>
String containing the path and name of the target file.
Example:
MMEM:STOR:LIST 'test'
Stores the current list evaluation results in the
test.dat
file.
Manual operation:
See
"Saving the Result Summary (Evaluation List) to a File"
on page 257
See
Managing settings and results
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...