
Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
866
User Manual 1178.8520.02 ─ 08
[SENSe:]POWer:ACHannel:PRESet:RLEVel
This command determines the ideal reference level for the current measurement.
This automatic routine makes sure that the that the signal power level does not over-
load the R&S
FSV/A or limit the dynamic range by too small a S/N ratio.
To determine the best reference level, the R&S
FSV/A aborts current measurements
and performs a series of test sweeps. After it has finished the test, it continues with the
actual measurement.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end. This is only possible in single sweep mode.
[SENSe:]POWer:TRACe
<TraceNumber>
This command selects the trace channel power measurements are performed on.
For the measurement to work, the corresponding trace has to be active.
Parameters:
<TraceNumber>
Range:
1 to 6
*RST:
1
Example:
POW:TRAC 2
Assigns the measurement to trace 2.
Manual operation:
See
12.5.3
Measuring the channel power and ACLR
All remote control commands specific to channel power or ACLR measurements are
described here.
See also
Chapter 12.5.2, "Configuring power measurements"
Managing measurement configurations
................................................................867
...................................................................................... 868
....................................................................................... 872
Selecting the reference channel
............................................................................874
General ACLR measurement settings
.................................................................. 881
Configuring MSR ACLR measurements
............................................................... 882
Performing an ACLR measurement
......................................................................912
Retrieving and analyzing measurement results
.................................................... 912
Programming examples for channel power measurements
..................................916
Configuring and performing measurements
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...