Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
879
User Manual 1178.8520.02 ─ 08
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute:STATe
<State>[,
<State>]
This command turns the absolute limit check for the alternate channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<ch>
1..n
Alternate channel number
Parameters:
<State>
ON | OFF | 1 | 0
Absolute limit check for lower alternate channel
*RST:
0
<State>
ON | OFF | 1 | 0
Absolute limit check for upper alternate channel
*RST:
0
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]
<LowerLimit>[,
<UpperLimit>]
This command defines the relative limit of the alternate channels. The reference value
for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S
FSV/A uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<ch>
1..n
Alternate channel number
Parameters:
<LowerLimit>
The limit of the lower alternate channel.
Range:
0 dB to 100 dB
*RST:
0 dB
Default unit: dB
Configuring and performing measurements
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...