Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1354
User Manual 1178.8520.02 ─ 08
INSTrument:COUPle:LLINes
<State>
This command turns synchronization of the limit lines between measurement channels
on and off.
Parameters:
<State>
ALL | NONE
ALL
Turns on synchronization.
NONE
Turns off synchronization.
*RST:
NONE
INSTrument:COUPle:MARKer
<State>
This command turns synchronization of the marker frequency between measurement
channels on and off.
Parameters:
<State>
ALL | NONE
ALL
Turns on synchronization.
NONE
Turns off synchronization.
*RST:
NONE
Example:
INST:COUP:MARK ALL
Synchronizes the receiver frequency and the marker frequency.
INSTrument:COUPle:PRESel
<State>
This command turns synchronization of the preselector state between measurement
channels on and off.
Parameters:
<State>
ALL | NONE
ALL
Turns on synchronization.
NONE
Turns off synchronization.
*RST:
ALL
Example:
INST:COUP:PRES ALL
Synchronizes the preselector configuration.