General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
755
User Manual 1178.8520.02 ─ 08
........................................................................................... 755
..........................................................................................756
NONE
Uses the current RF signal at the input, i.e. no calibration signal (default).
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
Calibration Frequency RF
Uses the internal calibration signal as the RF input signal.
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
DIAGnostic:SERVice:INPut:PULSed:CFRequency
Spectrum ← Calibration Frequency RF
Defines whether a broadband or narrowband calibration signal is sent to the RF input.
"Narrowband"
Used to calibrate the absolute level of the frontend at 64 MHz.
"Broadband"
Used to calibrate the IF filter.
Remote command:
DIAGnostic:SERVice:INPut:RF[:SPECtrum]
Service functions