Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
154
User Manual 1178.8520.02 ─ 08
c) By turning the rotary knob counterclockwise, reduce the trigger offset until the
burst edge is displayed in the center of the display, or enter
-50
µs
.
The R&S
FSV/A displays the rising edge of the GSM burst.
Figure 5-16: Rising edge of the GSM burst displayed with high time resolution
4. Move the falling edge of the burst to the center of the display. To do so, switch the
trigger "Slope" to "Falling" ([TRIG] > "Trigger/ Gate Config").
The R&S
FSV/A displays the falling edge of the GSM burst.
Figure 5-17: Falling edge of the GSM burst displayed with high time resolution