
General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
685
User Manual 1178.8520.02 ─ 08
During instrument start, the firmware checks whether the installed hardware is suppor-
ted. If not, an error message is displayed ("Wrong Firmware Version") and you are
asked to update the firmware. Until the firmware version is updated, self-alignment
fails.
If you start a self-alignment remotely and then select the "Local" softkey while the
alignment is still running, the instrument only returns to the manual operation state
after the alignment is completed.
During self-alignment, do not connect a signal to the RF input connector. Running a
self-alignment with a signal connected to the RF input can lead to false measurement
results.
Alignment results
The alignment results are displayed and contain the following information:
●
Date and time of last correction data record
●
Overall results of correction data record
●
List of performed alignment steps
The results are classified as follows:
PASSED
Calibration successful without any restrictions
CHECK
Deviation of correction value larger than expected, correction could however be per-
formed
FAILED
Deviations of correction value too large, no correction was possible. The found cor-
rection data is not applicable.
The results are available until the next self-alignment process is started or the instru-
ment is switched off.
Temperature check
During self-alignment, the instrument's frontend temperature is measured (if activated,
only after the instrument has warmed up completely, see
on page 689). This temperature is used as a reference for a
continuous temperature check during operation. If the current temperature deviates
from the stored self-alignment temperature by a certain degree, a warning is displayed
in the status bar. The warning indicates the resulting deviation in the measured power
levels. A status bit in the
STATUs:QUEStionable:TEMPerature
register indicates a
possible deviation. The current temperature of the frontend can be queried using a
remote command (see
SOURce<si>:TEMPerature:FRONtend
Touchscreen alignment
When the device is delivered, the touchscreen is initially calibrated. However, to ensure
that the touchscreen responds to the finger contact correctly, a touchscreen alignment
is required.
Alignment
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...