General instrument setup
R&S
®
FSVA3000/ R&S
®
FSV3000
759
User Manual 1178.8520.02 ─ 08
Relays Cycle Counter
The hardware relays built into the R&S
FSV/A may fail after a large number of switch-
ing cycles (see data sheet). The counter indicates how many switching cycles the indi-
vidual relays have performed since they were installed.
Remote command:
10.10
Synchronizing measurement channel configuration
Access
: [SETUP] > "Parameter Coupling"
Each of the applications of the R&S
FSV/A is usually treated as an independent entity
regarding their configuration: changing a setting in one measurement channel does not
automatically change the corresponding setting in another channel.
For example, changing the frequency in the spectrum application does not, by default,
change the frequency in the vector signal analysis (VSA) application.
However, sharing settings can be convenient for certain measurement tasks. The
R&S
FSV/A provides a tool to couple (or synchronize) selected parameters across
applications - the coupling manager.
The coupling managers allows you not only to couple parameters, but also markers
and lines accross applications.
..................................................................................759
User-defined parameter coupling
..........................................................................762
................................................................................................766
............................................................................768
Example for a user-defined parameter coupling
................................................... 770
10.10.1
General parameter coupling
Access
: [SETUP] > "Parameter Coupling" > "General"
The "General" tab of the coupling manager contains several parameters that you can
couple across all (active) measurement channels - if the channel supports the corre-
sponding parameter.
Synchronizing measurement channel configuration
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...