Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
876
User Manual 1178.8520.02 ─ 08
Parameters:
<LowerLimit>
The limit of the lower adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
<UpperLimit>
The limit of the upper adjacent channel.
Range:
-200 dBm to 200 dBm
*RST:
-200 dBm
Default unit: dBm
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute:STATe
<State>[, <State>]
This command turns the absolute limit check for the adjacent channels on and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
Absolute limit check for lower adjacent channel
*RST:
0
<State>
ON | OFF | 1 | 0
Absolute limit check for upper adjacent channel
*RST:
0
Manual operation:
See
See
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]
<LowerLimit>[,
<UpperLimit>]
This command defines the relative limit of the adjacent channels. The reference value
for the relative limit is the measured channel power.
If you have defined an absolute limit as well as a relative limit, the R&S
FSV/A uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
Configuring and performing measurements
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...