List Evaluation State
Activates or deactivates the list evaluation.
Remote command:
Show Peaks
If activated, all peaks that have been detected during an active list evaluation are
marked with blue squares in the diagram.
Remote command:
Margin
A margin functionality is not available for the limit check. However, you can define a
margin (=
threshold
) for the peak values to be displayed in the evaluation list. Only
peaks that exceed the margin value are displayed (also in the diagram, if activated).
Remote command:
CALCulate<n>:PEAKsearch:MARGin