Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1318
User Manual 1178.8520.02 ─ 08
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
SENS:CORR:FRES:USER:STAT
Manual operation:
See
[SENSe:]CORRection:FRESponse<si>:BASeband:USER:STORe
<FilePath>
[SENSe:]CORRection:FRESponse<si>:INPut<ip>:USER:STORe
<FilePath>
[SENSe:]CORRection:FRESponse<si>:USER:STORe
<FilePath>
Stores a saveset for a user-defined frequency response correction scenario. By
default, the saveset is stored in the
C:\R_S\INSTR\USER\FResponse
directory.
Suffix:
<si>
.
1..n
irrelevant
Setting parameters:
<FilePath>
string
Example:
SENS:CORR:FRES:USER:STOR 'FRes1'
Stores the current frequency response correction configuration
to a file named
C:\R_S\INSTR\USER\FResponse\FRes1.dfl
.
Usage:
Setting only
Manual operation:
See
[SENSe:]CORRection:FRESponse<si>:USER:VALid?
This command queries the validity of the user-defined correction settings.
Suffix:
<si>
.
1..n
irrelevant
Return values:
<Validity>
0 | 1
1
The setting is valid.
0
The setting is not valid.
Example:
SENS:CORR:FRES:USER:VAL?
Usage:
Query only
Manual operation:
See
Configuring the R&S
FSV/A
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...