Remote commands
R&S
®
FSVA3000/ R&S
®
FSV3000
1081
User Manual 1178.8520.02 ─ 08
Suffix:
<gen>
.
Parameters:
<Name>
<Generator name as string value>
*RST:
SMU02
Example:
//Select an external generator
SYST:COMM:RDEV:GEN:TYPE 'SMW06'
Manual operation:
See
SYSTem:COMMunicate:TCPip:RDEVice:GENerator<gen>:ADDRess
<Address>
Configures the TCP/IP address for the external generator.
Suffix:
<gen>
.
Parameters:
<Address>
TCP/IP address between 0.0.0.0 and 0.255.255.255
*RST:
0.0.0.0
Example:
SYST:COMM:TCP:RDEV:GEN:ADDR 130.094.122.195
Manual operation:
See
"TCPIP Address / Computer Name"
Source calibration
The following commands are required to activate the calibration functions of the exter-
nal tracking generator. However, they are only available if external generator control is
active (see
SOURce<si>:EXTernal<gen>[:STATe]
Useful commands for source calibration described elsewhere:
●
Chapter 12.10.3, "Working with transducers"
●
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:
RPOSition
Remote commands exclusive to source calibration:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
..........................1081
[SENSe:]CORRection:COLLect[:ACQuire]
......................................................................1082
..................................................................................... 1083
........................................................................................1083
.......................................................................................1083
[SENSe:]CORRection:TRANsducer:GENerate
................................................................ 1084
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RVALue
<Value>
This command defines the reference value assigned to the reference position in the
specified window. Separate reference values are maintained for the various displays.
Setting basic measurement parameters
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...