Getting started
R&S
®
FSVA3000/ R&S
®
FSV3000
73
User Manual 1178.8520.02 ─ 08
a) In the "Span" field of the "Frequency" dialog box, enter
20
MHz
.
b) Close the "Frequency" dialog box.
The display of the calibration signal is now improved. The maximum at the center
frequency (=calibration frequency) of 64
MHz becomes visible.
Figure 3-5: Calibration signal with optimized display settings
3.4.2
Displaying a spectrogram
In addition to the standard "level versus frequency" spectrum display, the R&S
FSV/A
also provides a spectrogram display of the measured data. A spectrogram shows how
the spectral density of a signal varies over time. The x-axis shows the frequency, the y-
axis shows the time. A third dimension, the power level, is indicated by different colors.
Thus you can see how the strength of the signal varies over time for different frequen-
cies.
1. Tap the "Overview" softkey to display the general configuration dialog box.
2. Tap the "Display Config" button.
The SmartGrid mode is activated, and the evaluation bar with the available evalua-
tion methods is displayed.