Measurements and results
R&S
®
FSVA3000/ R&S
®
FSV3000
193
User Manual 1178.8520.02 ─ 08
[SENSe:]POWer:ACHannel:FILTer[:STATe]:ALTernate<ch>
[SENSe:]POWer:ACHannel:FILTer[:STATe]:UACHannel
[SENSe:]POWer:ACHannel:FILTer[:STATe]:UALTernate<ch>
Alpha value:
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ACHannel
[SENSe:]POWer:ACHannel:FILTer:ALPHa:ALTernate<ch>
[SENSe:]POWer:ACHannel:FILTer:ALPHa:UACHannel
[SENSe:]POWer:ACHannel:FILTer:ALPHa:UALTernate<ch>
Limit Checking ← Adjacent Channel Definition
During an ACLR measurement, the power values can be checked whether they
exceed user-defined or standard-defined limits. A relative or absolute limit can be
defined, or both, for each individual adjacent channel. Both limit types are considered,
regardless whether the measured levels are absolute or relative values. The check of
both limit values can be activated independently. If any active limit value is exceeded,
the measured value is displayed in red and marked by a preceding asterisk in the
result table.
Note that in addition to activating limit checking for individual channels, limit checking
must also be activated globally for the MSR ACLR measurement (see
Remote command:
CALCulate<n>:LIMit<li>:ACPower[:STATe]
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute:STATe
CALCulate<n>:LIMit<li>:ACPower:ACHannel:ABSolute
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]:STATe
CALCulate<n>:LIMit<li>:ACPower:ACHannel[:RELative]
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute:STATe
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>:ABSolute
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]:STATe
CALCulate<n>:LIMit<li>:ACPower:ALTernate<ch>[:RELative]
CALCulate<n>:LIMit<li>:ACPower:ACHannel:RESult?
5.3.5.4
MSR gap channel setup
Access
: "Overview" > "Select Measurement" > "Channel Power ACLR" > "CP / ACLR
Standard" > "Standard": "Multi-Standard Radio" > "CP / ACLR Config" > "Gap Chan-
nels" tab
The "Gap Channels" tab provides all the channel settings to configure gap channels in
MSR ACLR measurements.
Channel power and adjacent-channel power (ACLR) measurement
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...