Getting started
R&S
®
FSVA3000/ R&S
®
FSV3000
77
User Manual 1178.8520.02 ─ 08
Figure 3-9: Frequency spectrum of the calibration signal with a larger span
4. Repeat the previous steps to activate a third Spectrum window.
Change the frequency range for this spectrum display:
In the "Frequency" dialog box, set the
center frequency
to
64
MHz
and tap "Zero
Span".
As the calibration signal does not vary over time, the level versus time diagram dis-
plays a straight line.
Trying out the instrument