Common measurement settings
R&S
®
FSVA3000/ R&S
®
FSV3000
462
User Manual 1178.8520.02 ─ 08
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RPOSition
Auto Scale Once
Automatically determines the optimal range and reference level position to be dis-
played for the current measurement settings.
The display is only set once; it is not adapted further if the measurement settings are
changed again.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
Scaling
Defines the scaling method for the y-axis.
For more information, see
"Logarithmic"
Logarithmic scaling (only available for logarithmic units - dB..., and A,
V, Watt)
"Linear with
Unit"
Linear scaling in the unit of the measured signal
"Linear Per-
cent"
Linear scaling in percentages from 0 to 100
"Absolute"
The labeling of the level lines refers to the absolute value of the refer-
ence level (not available for "Linear Percent")
"Relative"
The scaling is in dB, relative to the reference level (only available for
logarithmic units - dB...). The upper line of the grid (reference level) is
always at 0 dB.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
6.4.4
How to optimize the amplitude display
This section gives you some advice on how to optimize the display of the measured
signal amplitudes depending on the required evaluation.
1. Perform a measurement with the default settings to get an impression of the values
to be expected.
2. Use the "Auto Level" function ([AUTO] menu) to optimize the reference level.
3. Use the "AF Auto Scale" function ([AUTO] menu) to optimize the scaling.
4.
To determine a precise level at a specific point in the signal:
Amplitude and vertical axis configuration
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...