Data management
R&S
®
FSVA3000/ R&S
®
FSV3000
641
User Manual 1178.8520.02 ─ 08
Note:
Saving instrument settings in secure user mode.
Settings that are saved via Quick Save in secure user mode are only available during
the current session. As soon as the power is switched off on the R&S
FSV/A, the data
is cleared (see
Chapter 3.2.15, "Protecting data using the secure user mode"
Rename ← QuickSave 1 / ... / QuickSave 10
Displays an input field to rename the save set, if write protection is disabled.
Write Protection ← QuickSave 1 / ... / QuickSave 10
Enables or disables write protection for the save set. If enabled, the save set cannot be
renamed or overwritten.
Storage Type (Save only)
Defines which type of settings are stored in the save set.
"Instrument
with all Chan-
nels"
The instrument settings for all currently active "Channel"s are stored.
"Current Chan-
nel"
Only the instrument settings for the currently selected measurement
"Channel"s are stored.
Recall
Restores the instrument settings as saved in the selected settings file. If the settings
file contains settings for a specific "Channel" only, a new channel with the stored set-
tings is activated, otherwise all "Channel"s and instrument settings are overwritten with
the stored settings.
Note:
After you use the "Recall" function, the history of previous actions is deleted, i.e.
any actions performed previously cannot be undone or redone using the [UNDO/
REDO] keys.
Remote command:
9.3.2
Configurable storage and recall
The more sophisticated storage and recall functions allow you to define which settings
are stored, and where the settings file is stored to. Any settings file can be selected for
recall.
...............................................................................642
................................................................................ 642
............................................................................................645
9.3.2.1
Stored data types
The following types of data can be stored to and loaded from files via the "Save" dialog
box on the R&S
FSV/A:
Storing and recalling instrument settings and measurement data
Содержание FSV3000
Страница 1: ...R S FSVA3000 R S FSV3000 Signal and Spectrum Analyzer User Manual 1178852002 Version 08 ÜãD2 ...
Страница 32: ...Contents R S FSVA3000 R S FSV3000 32 User Manual 1178 8520 02 08 ...
Страница 287: ...Measurements and results R S FSVA3000 R S FSV3000 287 User Manual 1178 8520 02 08 Statistical measurements APD CCDF ...
Страница 673: ...Data management R S FSVA3000 R S FSV3000 673 User Manual 1178 8520 02 08 Working with test reports ...