Remote control commands
R&S
®
NRPM
160
User Manual 1425.8663.02 ─ 07
HOLD
Triggers the measurement with the command
.
IMMediate
Starts the measurement immediately.
INTernal | INT1 | INT2 | INT3
Determines to consider the RF signal level at the specified
antenna (1, 2 or 3) for launching a measurement. The
parameter
INTernal
is a synonym for
INT1 |
INTernal1
.
The measurement starts when the signal exceeds
(
) or drops below (
has elapsed. Similar to trigger
source
EXT
, waiting for a trigger event can also be skipped
.
*RST:
IMMediate
Manual operation:
TRIGger:SYNC:PORT
<sync_port>
Selects the external connection for the sensor module's sync output, see also
.
Parameters:
<sync_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
TRIGger:SYNC:STATe
<state>
Usually used in combination with
If activated, blocks the external trigger bus as long as the sensor modules
remains in the measurement state.
This function makes sure, that a new measurement only starts after all sensor
modules have completed their last measurement.
Configuring the trigger