Remote control commands
R&S
®
NRPM
158
User Manual 1425.8663.02 ─ 07
You can assign either the port
EXTernal[1]
or
EXTernal2
for the trigger
sender. If you want to trigger the sender sensor module externally with its own
trigger signal, assign the trigger source to the other port accordingly, e.g:
TRIGger:SENDer:PORT EXT1
TRIGger:SOURce
EXT2
TRIGger:SENDer:
STATe
ON
or
TRIGger:SENDer:PORT EXT2
TRIGger:SOURce
EXT1
TRIGger:SENDer:
STATe
ON
Parameters:
<sender_port>
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
TRIGger:SENDer:STATe
<state>
Activates the trigger sender mode of the sensor module. In this state, the sensor
module can output a digital trigger signal in synchronization with its own trigger
event.
If activated, select the output port the trigger signal using
.
Typically, the trigger sender uses its internal trigger source. But you can also trig-
ger the trigger sender externally, since the R&S
NRPM provides two external trig-
ger connectors. If you trigger the sender externally, use
EXTernal1
as external
trigger input port (trigger source) and
EXTernal2
as trigger sender output port or
vice versa.
Parameters:
<state>
*RST:
OFF
Configuring the trigger