Remote control commands
R&S
®
NRPM
120
User Manual 1425.8663.02 ─ 07
Measurement start
Depending on whether you want to measure continuously or want to execute one
dedicated measurement, select the corresponding command:
●
Starts a new measurement cycle automatically after the previous one has
been completed.
●
Starts a single measurement only. Depending on the selected measurement
function, a
single measurement
can lead to an array of results, e.g., in trace
mode.
See
Trigger sources
The possible trigger conditions and the execution of a trigger depend on the
selected trigger source.
If the signal power exceeds or falls below a reference level set by the trigger
level, the measurement starts after the defined delay time. Waiting for a trigger
event can be skipped by selecting
Trigger
source
Description
Remote commands to initiate the
measurement unconditionally
Hold
Triggered by the remote command.
Immediate
Measures immediately, does not
wait for trigger condition.
-
Internal
Uses the input signal as trigger sig-
nal.
External
1
Uses the digital input signal sup-
plied using a differential pair in the
8-pin sensor module cable.
External
2
Uses the digital input signal sup-
plied at the SMB connector.
Bus
Triggered by the remote command.
Controlling the measurement