Remote control commands
R&S
®
NRPM
157
User Manual 1425.8663.02 ─ 07
TRIGger:IMMediate
Triggers a generic trigger event that causes the sensor module to exit the
WAIT_FOR_TRIGGER state immediately, irrespective of the trigger source and
the trigger delay and start the measurement. The command is the only means of
starting a measurement when the trigger source is set to HOLD. Only one mea-
surement cycle is executed irrespective of the averaging factor.
Usage:
Event
TRIGger:LEVel
<level>
Sets the trigger threshold for internal triggering derived from the test signal.
This setting is effective only for the internal
Parameters:
<level>
Range:
1.0e-7 to 200.0e-3
*RST:
1.0e-3
Default unit: W
Manual operation:
TRIGger:LEVel:UNIT
<unit>
Sets the unit of the trigger level if this value is entered without a unit.
See
Parameters:
<unit>
DBM | W | DBUV
*RST:
W
Example:
TRIGger:LEVel:UNIT W
Manual operation:
TRIGger:SENDer:PORT
<sender_port>
Selects the port where the R&S
NRPM provides its internal trigger signal at the
output, provided it is trigger sender (see
Configuring the trigger