To set up the connection using the sensor hub (
):
1. Connect the R&S
NRP
‑
Z5 to the computer.
2. Connect the NRP-ZK6 to the power sensor, see
3. Connect the USB connector of the NRP-ZK6 to the R&S
NRP
‑
Z5 sensor hub.
4. If necessary, connect further sensor modules to the sensor hub accordingly.
5. Connect the external power supply unit to the R&S
NRP
‑
Z5 and to the mains.
6. Establish the connection of the antenna modules as described in
ter 6.3.1, "Setting up the RF frontend"
7. On the computer, start a software application to execute the measurement,
Chapter 7, "Performing measurements"
6.3.4
Using the LAN connection
Requires the R&S
NRPM3N LAN sensor module with networking capabili-
ties.
The Ethernet interface of a R&S
NRPM3N requires PoE (power over Ether-
Electromagnetic interference (EMI) can affect the measurement results. To
avoid any impact, use at least category 5 cables or better.
Risk of sensor damage
Use only PoE power sourcing equipment (PSE) according to IEEE stand-
ards 802.3af or IEEE 802.3at.
Otherwise your power sensor can get damaged.