10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-24
No.
Message
Interpretation
Operator action
1009 Unable to
calculate oxygen
parameter
Can be shown on a result if
unable to calculate oxygen
due to an unexpected system
error.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1010 Oxi data
collection error
Oxi hardware problem
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1011 Oxi has no Blank
Cal
Missing Blank Cal. Not
necessarily a hardware error.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
Removal conditioning:
- Successful Blank
calibration
1012 Oxi has no
sample spectrum
The system has not made a
sample measurement yet, or
there is a hardware problem.
- Repeat the
measurement.
- Restart the analyzer.
- Contact Radiometer
service representative.
1013 Oxi data
collection error
Oxi hardware error
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
1014 Oxi Blank Cal.
intensity too high
The spectrometer received
too high light intensity during
Blank Cal.
- Check solution pack.
During Oxi Blank
calibration, the cuvette
must be filled with
liquid.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
1015 Oxi sample
intensity too high
The spectrometer received
too high light intensity during
sample measurement.
- Check solution pack.
During Oxi Blank
calibration, the cuvette
must be filled with
liquid.
- Perform a calibration.
- Repeat the sample
measurement.
1016 Oxi Blank Cal.
intensity too low
The spectrometer received
too low light intensity during
Blank Cal.
- Perform a calibration.
- Restart the analyzer.
- Contact Radiometer
service representative.
1017 Oxi sample
intensity too low
The spectrometer received
too low light intensity during
sample measurement.
- Restart the analyzer.
- Perform a calibration.
- Contact Radiometer
service representative.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...