1. Setup
ABL90 FLEX reference manual
1-50
Barometer setup
In this program you can adjust the automatic barometer in accordance with the
reference barometer in your laboratory.
Step Action
1.
Key in the desired pressure value on the keypad.
2.
Confirm with Enter. The value will be shown in the "Measured
adjusted" box.
Maximum accepted correction is ±19 mmHg (i.e. the difference
between the "Measured unadjusted" and "Measured adjusted"
settings).
Barometer pressure limits are 450-800 mmHg, or 60.0-106.7 kPa,
or 450-800 Torr.
The units are selected in the Setup program: Units.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...