3. Wet section
ABL90 FLEX reference manual
3-6
Rinse process
After a measurement is complete a rinse is performed. The rinse process is the
same, no matter what kind of measurement (patient measurement, QC and
calibration) is performed. The following table describes this rinse process.
Stage Description
1
After the measurement is complete the first part of the rinse is
performed with a mixture of solutions and air.
2.
The next part of the rinse is performed with a mixture of solutions
and gas.
3.
Thereafter the wet section is checked. The system is filled with gas
to equilibrate the measuring chambers.
4.
The entire measuring path is filled with CAL 1 (rinse) solution. The
calibration status is reestablished and the device is now ready for a
new measurement.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...