ABL90 FLEX reference manual
10. Troubleshooting, updated for software version 2.3
10-7
No.
Message
Interpretation
Operator action
214
Automatic
backup
succeeded
The scheduled automatic
backup was completed
successfully.
- No action required.
216
General printer
error
A printer problem has
occurred, e.g. the paper is
jammed
- Check printer paper.
Clear any jam.
- Power down and
restart the analyzer.
- Contact Radiometer
service representative.
217
Replacement:
The message is used in the
Activity Log to indicate a
performed replacement.
- No action required.
290
Warning: SHb
detected
FSHb detected in the range
of 1-10 %.
- No action required. For
information only.
291
SHb too high
Detected FSHb is greater
than 10%. Measurement
accuracy is affected.
- Repeat the
measurement.
292
Turbidity too
high
Turbidity is greater than 5
%: too high for reliable
measurements.
- Hyperlipemic sample;
decrease the lipemic
content by e.g.
centrifuge or extraction.
- Perform the
measurement on a
blood sample from a
healthy donor.
- Contact Radiometer
service representative.
293
Oxi compensated
for HbF
OXI parameters have been
HbF compensated. Parameter
FHbF may be shown or not
shown.
- No action required. For
information only.
329
QC expiration
date exceeded
The quality control
measurement was performed
on an expired control
solution.
- Discontinue the use of
the lot and set up a
valid lot for the control
solution.
331
No sample
detected during
sample aspiration
No sample detected in
sensor.
Measurement is aborted.
- Ensure that adequate
sample volume is used.
- Check the sample for
clots.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...