ABL90 FLEX reference manual
Appendix – Quality control
I-9
c
tHb g/dL
0.06
F
HbF %
0.4
Other Hb derivatives %
0.1
Statistics factor and statistics range
Normal statistical variation implies that 95.5 % of all quality control results
obtained on a well-functioning analyzer falls within
2 SD range, and 99.7 %
falls within
3 SD range.
In order to include all results from a well-functioning analyzer, a statistics factor
of 1.5 (default) is used to expand the control ranges. This also ensures that user
control ranges do not become too narrow over time.
Using the recommended statistics factor of 1.5 will have the following effect:
User control ranges (2 SD): The statistics range will correspond to
X
3 SD.
Insert control ranges:
The statistics range will correspond to 1.5
insert control range.
All results outside the statistics range will be excluded from the statistics and
marked accordingly.
The control range is: pH low = 6.986 and pH high = 7.016
To calculate the statistics range, do the following:
Step Action
1.
Calculate the mean value:
X = (6.986 + 7.016)/2 = 7.001.
2.
Calculate the 2 SD: pH high
X = 7.016
7.001 = 0.015.
3.
Calculate the 3 SD: (0.015
3)/2 = 0.0225 = 0.023.
4.
The statistics range will then be:
pH low = 7.001
0.023 = 6.978.
pH high = 7.001 + 0.023 = 7.024.
Definitions
Example
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...