10. Troubleshooting, updated for software version 2.3
ABL90 FLEX reference manual
10-12
No.
Message
Interpretation
Operator action
589
Measured QC
value higher than
statistical range
The parameter value is
above the upper limit of the
user-defined statistical
range. Measurement not
included into statistics.
- Verify the procedure
and repeat the
measurement.
- Refer to the ABL90
FLEX reference manual
for details on the
evaluation of the
results.
593
Insufficient
sample
Sample volume is too small
for the selected measuring
mode. Affected parameters
will be marked with "?".
- Repeat the
measurement, ensuring
sufficient sample
volume.
- Contact Radiometer
service representative.
595
Liquid sensor
calibration error
One or more of the liquid
sensors failed calibration.
- Repeat the liquid
sensor calibration.
- Check solution pack
status and replace, if
necessary.
- Contact Radiometer
service representative.
606
Cal expired (pH) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful calibration.
608
Cal expired
(pCO
2
)
Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
609
Cal expired (pO
2
) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
610
Cal expired (K)
Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
611
Cal expired (Na) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
612
Cal expired (Ca) Too long time passed since
the last successful calibration
of the parameter. Parameter
measurement values are
reported as ".....".
- Perform a calibration.
Removal condition:
- Successful 2-point
calibration.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...