5. Sensors and optical system
ABL90 FLEX reference manual
5-42
The figure below shows three spectra; pure O
2
Hb, pure HHb in a low
concentration, a spectrum of 92 % oxygenated hemoglobin obtained by adding
the spectra of O
2
Hb and HHb. The additivity of absorption and the continuity of
the spectra can be seen.
480
500
520
540
560
580
600
620
640
660
680
O
2
Hb (9.2 mmol/L)
HHb (0.8 mmol/L)
92 % oxygenated hemoglobin (i.e., 92 % O
2
Hb + 8 % HHb)
Wavelength/nm
Absorption
Example of the spectrum obtained from unconjugated bilirubin at a
concentration of 200 µmol/L.
The spectrum of conjugated bilirubin is slightly different.
In the spectrum taken of a sample, the absorption recorded at each wavelength
contains contributions from each of the compounds in the sample. The task then
is to determine the magnitude of that contribution and thereby the
concentration of each compound in the sample.
The concentrations are determined using the following equation:
138
y
total
1
K
n
n
y
n
c
A
where:
n
y
K
= a constant specific to compound y at wavelength
n
.
Spectrum
examples
200u m o l/L U n co n ju g ated B iliru b in in P lasm a
0
0.02
0.04
0.06
0.08
0.1
470
520
570
620
670
nm
Ab
s
o
rb
a
n
ce
Determining
concentrations
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...