ABL90 FLEX reference manual
1. Setup
1-27
Step Action
1.
Press Next slot to display the desired slot and then press Edit.
2.
Select the parameter to be edited, using Next param. or Prev
param.
3.
Press Update to change the range to the one shown in the "Lot to
Date range (2 SD)" box (if available).
4.
Press the check button to activate or deactivate the Fixed SD (i.e. a
minimum allowed control range is defined by setting a Fixed SD).
To change the SD value, touch the "SD" field to highlight it and
enter the value, using the keypad. Confirm the entry with Enter.
5.
Highlight the limit by touching it on the screen and enter your own
value(s), using the keypad. Confirm with Enter.
6.
Repeat the procedure for other parameters in the same manner.
NOTICES:
See I Appendix - Quality control in this manual for detailed
information on statistics and its parameters
"Lot-to-date range (2 SD)" is the range calculated over the
course of the lot, represented mathematically by the mean
value
2 SD; this is the range within which 95 % of the
measurements are found.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...