ABL90 FLEX reference manual
Appendix – Quality control
I-5
Term
Explanation
At Radiometer
Assigned value The center value of a
control range.
There is one assigned
value for each
parameter for each
type of analyzer.
When a quality control system is
developed, it is tested on 10 well-
functioning analyzers of the type for
which the quality control system
has been designed. Measurements
on level 1 – level 2 – level 3 – level
4 are performed two to five times
during 24 hours. These
measurements are repeated for 1–
4 weeks in order to give 30 – 50
measurements per analyzer.
From true
value to
assigned value
The true value of a parameter is
correlated to the assigned value by
an algorithm for a particular type of
analyzer and QC product. This is a
one-time-only act.
Insert control
ranges
The interval within
which a quality control
result of a well-
functioning analyzer
should fall with at least
95 % probability.
The insert control ranges are
established by being centered
around each parameter's assigned
value.
The width of the insert control
ranges is determined using an
uncertainty budget comprising
contributions from the analyzer, the
calibration solutions and the QCs.
The uncertainty budget guarantees
that only relevant contributions to
the width of the insert control
ranges are included.
User control
ranges
(analyzer-
specific)
A range established by
the user based on
results obtained on
one analyzer.
The inserts have control ranges for all analyzers for which the quality control
solution can be used. By using the data codes on the inserts, the information
about the type and level of the quality control solution and control ranges for all
parameters are transferred to the analyzer. The analyzer is then able to
recognize the quality control solutions.
NOTICE: The insert control ranges are determined at sea level. At high altitudes
the control ranges from the insert should be corrected as described later in this
appendix.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...