ABL90 FLEX reference manual
1. Setup
1-31
Activating Westgard Rules for a specific parameter:
Step Action
1.
Display the desired slot, using Next slot and press Edit.
2.
Display the desired parameter using Next param or Prev param.
3.
Activate the desired Westgard Rule(s) by pressing the
corresponding check button.
(All future quality control data for the given slot/parameter will be
evaluated according to the selected Westgard Rule(s).)
4.
Select Westgard Rules for other parameters or levels in the same
manner.
5.
Press Back to return to the Westgard Rules setup screen.
Selecting/deselecting all Westgard Rules:
Step Action
1.
Display the desired slot, using Next slot.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...