ABL90 FLEX reference manual
5. Sensors and optical system
5-35
Measurement – metabolites
The glucose or lactate concentration in a sample is calculated from the following
equation, using the difference between the current in the sample and the
extrapolated zero current from the rinse solution:
sample
0
I
I
c
S
The measured value is applied a linear correction:
displayed
1
2
c
k c k
NOTICE
:
cLac is compensated for the dependence of the ionic composition by
using the measured electrolyte values before the linear correction is applied. If
the electrolytes are not measured, default values are used.
The following parameters are range checked:
Sensitivity
Sensitivity drift
Sensor response stability
The sensor response stability is defined as the standard deviation of the last 5
updatings of the response for CAL 1 (rinse).
For CAL 3, it is defined as the standard deviation of a linear regression for the
last 5 samples, normalized with the signal magnitude.
Measurement
Checks
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...