ABL90 FLEX reference manual
7. Performance characteristics
7-17
Performance test results – FHHb
0%: whole blood is tonometered with a gas mixture containing 94.4% N
2
and 5.6% CO
2
+ dithionite.
Setup:
Adult samples. HbF correction is not activated.
FHHb
(%)
ctHb
(g/dL)
Bias
Sec.ref
S
0
S
x
CV
X
% TE
A
TE
A
(%)
0.0 15 0.07 0.10
0.28 - 0.62 -
10.0 15 0.08 0.18 0.35 3.5 0.77 7.7
25.0 15 0.05 0.30 0.48 1.9 0.99 4.0
35.0 15 0.08 0.27 0.50 1.4 1.06 3.0
50.0 15 0.11 0.26 0.57 1.1 1.23 2.5
100.0 15 -0.14 0.16 0.40 0.4 0.92 0.9
0.0 7 -0.45
0.13
0.36 - 1.16 -
0.0 25 0.53 0.09
0.26 - 1.04 -
Performance test results – FHbF
The reference method is based on Cation Exchange HPLC. The method is described in
[Ref. 15]. The method is performed by the Hæmatology Laboratory at Herlev Hospital,
Denmark.
Setup:
Mixed adult and Fetal samples. HbF correction enabled for all levels.
FHbF
(%)
ctHb
(g/dL)
Bias
Prim.ref
*)
S
0
S
X
CV
X
% TE
A
TE
A
(%)
0 15 1.6 1.8
3.0 - 7.48 -
5 15 3.2 1.8
3.3 66
9.67
193
10 15 0.9 1.7 3.2 32 7.17
71.7
20 15 1.1 1.8 3.6 18 8.16
40.8
30 15 -1.7 1.8 4.2 14 9.93
33.1
50 15 -3.1 1.6 3.9 7.8
10.74
21.5
80 15 -4.2 1.8 4.1 5.1
12.24
15.3
*) ABL735 corrected to HPLC through:
F
HbF(ABL735),corr = FHbF (ABL735),meas – 0.9* ctHb + 11.7
Reference
method
Bias
Sec.ref
and
Repeatability –
blood samples
Reference
method
Bias
Prim.ref
and
Repeatability –
blood samples
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...