7. Performance characteristics
ABL90 FLEX reference manual
7-6
Test conditions
Test conditions to determine bias
Prim.ref
, bias
Sec.ref
, repeatability, reproducibility and total
variation for pH, pCO
2
, pO
2
, cCa
2+
, cCl
–
, cK
+
, cNa
+
, cGlu, cLac, ctHb, sO
2
, FO
2
Hb,
F
COHb, FMetHb, FHHb, ctBil and FHbF were as follows:
Item
Description
Reference analyzers
Five ABL735 analyzers with AutoCheck module were
used as a reference. The capillary mode was used
for pCO
2
and pO
2
, and the syringe mode for all the
other parameters.
Primary/secondary
reference methods
As specified for each parameter earlier in this
chapter.
Analyzers and test
modes
8-10 ABL90 FLEX analyzers were tested in syringe
and capillary mode.
Blood samples
Heparinized blood samples from healthy, voluntary
donors.
The blood is prepared to obtain different
concentration levels of each measured parameter.
Blood measurements
Measurements on every parameter are performed
on all analyzers, with 5 measurements on every
sample of each run, repeated for 3 days.
The measurements were performed by different
operators.
Solution pack
All calibration solutions and gases used for the tests
are traceable to Primary Reference Standards.
Traceability certificates for the ABL90 FLEX
calibration solutions and gases are found at the end
of chapter 9: Solutions.
Experimental
conditions
Ambient temperature: 22-25 °C
Relative humidity: 30-50 %.
Barometric pressure: 730 - 780 mmHg
NOTICES:
The solutions used in the performance tests are those
recommended by Radiometer. Performances using other
solutions cannot be guaranteed.
The performance tests are performed under conditions where
the analyzers are not influenced by electromagnetic fields.
Содержание FLEX ABL90
Страница 1: ...ABL90 FLEX reference manual ...
Страница 2: ......
Страница 3: ...ABL90 FLEX reference manual ...
Страница 4: ......
Страница 124: ...3 Wet section ABL90 FLEX reference manual 3 10 ...
Страница 125: ...4 Electronics General information 4 2 Electronic boards and components 4 3 ...
Страница 216: ...7 Performance characteristics ABL90 FLEX reference manual 7 30 ...
Страница 263: ...9 Solutions General information 9 2 Solutions 9 3 Certificate of traceability 9 5 ...
Страница 267: ...ABL90 FLEX reference manual 9 Solutions 9 5 Certificate of traceability ...
Страница 268: ...9 Solutions ABL90 FLEX reference manual 9 6 ...
Страница 269: ...10 1 10 Messages List of analyzer messages 10 2 ...
Страница 310: ...10 Troubleshooting updated for software version 2 3 ABL90 FLEX reference manual 10 42 ...
Страница 326: ...Appendix Quality control ABL90 FLEX reference manual I 16 ...
Страница 327: ...II Appendix Traceability to the primary standards at Radiometer ...
Страница 334: ...Appendix Traceability to the primary standards at Radiometer ABL90 FLEX reference manual II 8 ...
Страница 346: ...Date of issue ABL90 FLEX reference manual ...